Explosives trace detector

Explosive trace detectors (ETD) are explosive detection equipment able to detect explosives of small magnitude. The detection is accomplished by sampling non-visible "trace" amounts of particulates. Devices similar to ETDs are also used to detect narcotics. The equipment is used mainly in airports and other vulnerable areas considered susceptible to acts of unlawful interference.

Characteristics

Sensitivity

Detection limit is defined as the lowest amount of explosive matter a detector can detect reliably. It is expressed in terms of nano-grams (ng), pico-grams (pg) or femto-grams (fg) with fg being better than pg better than ng. It can also be expressed in terms of parts per billion (ppb), parts per trillion (ppt) or parts per quadrillion (ppq).

Sensitivity is important because most explosives have a low vapor pressure . The detector with the highest sensitivity is the best in detecting vapors of explosives reliably.

Light weight

Portable explosive detectors need to be as light weight as possible to allow users to not fatigue when holding them. Also, light weight detectors can easily be placed on top of robots.

Size

Portable explosive detectors need to be as small as possible to allow for sensing of explosives in hard to reach places like under a car or inside a trash bin.

Cold start up time and analysis time

The start up time for any trace detector is the time required by the detector to reach the optimized temperature for detection of contraband substances.

Technologies

Colorimetrics

The use of colorimetric test kits for explosive detection is one of the oldest, simplest, and most widely used methods for the detection of explosives. Colorimetric detection of explosives involves applying a chemical reagent to an unknown material or sample and observing a color reaction. Common color reactions are known and indicate to the user if there is an explosive material present and in many cases the group of explosive from which the material is derived. The major groups of explosives are nitroaromatic explosives, nitrate ester and nitramine explosives, improvised explosives not containing nitro groups which includes inorganic nitrate based explosives, chlorate based explosives, and peroxide based explosives.

Immunoassay

Explosive detection immunoassay is a sensitive, antibody-based analytical method used to identify trace amounts of energetic compounds like TNT, RDX, PETN, and HMX in environmental samples, forensic scenes, or security screening. It relies on specific antigen-antibody binding reactions translated into optical or colorimetric signals.

Ion mobility spectrometry

Explosive detection using ion mobility spectrometry (IMS) is based on velocities of ions in a uniform electric field. There are some variant to IMS such as Ion trap mobility spectrometry (ITMS) or Non-linear dependence on ion mobility (NLDM) which are based on IMS principle. The sensitivity of devices using this technology is limited to pg levels. The technology also requires the ionization of sample explosives which is accomplished by a radioactive source such as nickel-63 or americium-241. This technology is found in most commercially available explosive detectors such as the GE VaporTracer, Smith Sabre 4000 and Russian built MO-2M and MO-8.[citation needed] The presence of radioactive materials in these equipments cause regulatory hassles and requires special permissions at customs ports. These detectors cannot be field serviced and may pose radiation hazard to the operator if the casing of the detector cracks due to mishandling. Bi-yearly[clarification needed] checks are mandatory on such equipment in most countries by regulating agencies to ensure that there are no radiation leaks. Disposal of these equipments is also controlled owing to the high half-life of the radioactive material used.

Electrospray ionization, mobility analysis (DMA) and tandem mass spectrometry (MS/MS) is used by SEDET (Sociedad Europea de Detección) for the “Air Cargo Explosive Screener (ACES)”, targeted to aviation cargo containers currently under development in Spain.[citation needed]

Nuclear quadrupole resonance

Nuclear quadrupole resonance (NQR) is a radio-frequency spectroscopic technique used to detect and identify explosive materials—such as RDX, TNT, and HMX—by targeting nitrogen-14 (N14) nuclei without needing an external magnetic field.

Lab-on-a-chip

A lab-on-a-chip (LOC) platform combining immunoassays with microfluidics enables rapid, portable, and ultra-sensitive detection of trace explosives (such as TNT, DNT, and related energetic materials) or associated threat agents.

Thermo redox

This technology is based on decomposition of explosive substance followed by the reduction of the nitro groups. Most military grade explosives are nitro compounds and have an abundance of NO2 groups on them. Explosive vapors are pulled into an adsorber at a high rate and then pyrolized. The presence of nitro groups in the pyrolized products is then detected. This technology has significantly more false alarms because many other harmless compounds also have an abundance of nitro groups. For example, most fertilizers have nitro groups which are falsely identified as explosives, and the sensitivity of this technology is also fairly low. A popular detector using this technology is Scintrex Trace EVD 3000.

Laser-induced breakdown spectroscopy

Laser-induced breakdown spectroscopy (LIBS) is a rapid, in-situ analytical technique used to detect and identify chemical explosives through laser-generated microplasmas.

Chemiluminescence

This technology is based on the luminescence of specific compounds when they bind to explosive particles. Its sensitivity can reach the nanogram (ppb) level. It can be applied in non-electronic formats, such as sprays and test papers, as well as in electronic devices.

miniEXPLONIX Pro - swiss made fast & reliable explosive trace detector for field use with simple control

A key advantage of chemiluminescence-based electronic devices is their high reliability and resistance to interference from substances such as perfumes, dirt, and human sweat, as well as their ability to operate effectively under challenging environmental conditions, including variations in temperature and humidity. Therefore they are very capable in the field. These devices does not uses radioactive ionizing source.

These features are enabled by IRSSIL technology developed by RS DYNAMICS, powering the compact yet fast and capable miniEXPLONIX Pro detector, as well as the advanced EXPLONIX 2 analyzer.

Terahertz spectroscopy

THz radiation (0.3–10 THz) penetrates nonmetallic and nonpolar barriers like clothing, paper, and cardboard, interacting with unique low-energy intramolecular and phonon vibrational modes that act as spectral fingerprints for explosives.

Gamma-ray spectroscopy

Inertial Electrostatic Confinement (IEC) fusion devices detect explosives by using compact, tabletop D-D (deuterium-deuterium) fusion reactions to emit 2.45 MeV fast neutrons.

Amplifying fluorescent polymer

Amplifying fluorescent polymer (AFP) is a promising new technology and is based on synthesized polymers which bind to explosive molecules and give an amplified signal upon detection. When compounds that are not polymers are utilized for such purpose, the quenching of the fluorescence by the traces of explosives is not detectable. When amplifying fluorescent polymer in thin films absorbs a photon of light, excited state polymers (excitons) are able to migrate along the polymer backbone and between the adjacent polymer films. These sensors were originally made in order to detect trinitrotoluene. In AFP, binding of one TNT molecule results in quenching of fluorescence significantly due to the conjugated structure of the polymers. It has been reported that in practice the polymers result in 100-1000 fold increase of amplification of the quenching response.

"During its excited state lifetime, the exciton propagates by a random walk through a finite volume of the polymer film." Once TNT, or any other electron-deficient (i.e., electron accepting) molecule comes in contact with the polymer, a so-called low-energy ‘trap’ forms. "If the exciton migrates to the site of the bound electron-deficient molecule before transitioning back to the ground state, the exciton will be trapped (a non-radioactive process), and no fluorescence will be observed from the excitation event. Since the exciton samples many potential analyte binding sites during its excited state lifetime, the probability that the exciton will sample an occupied ‘receptor’ site and be quenched is greatly increased."

An example of an amplifying fluorescent polylmer

The explosive trace detectors utilizing AFPs, known as Fido Explosives Detectors, were originally developed under the Defense Advanced Research Projects Agency (DARPA) Dog’s Nose program and is now produced by FLIR Systems. The current generation, provides broad-band trace explosive detection and weighs less than 3 lbs. The sensitivity is in the order of femtogram (1 × 10−15 grams). This is the only such technology in the field that can achieve such sensitivity.

Mass spectrometry

Recently, mass spectrometry (MS) has emerged as another ETD technology. Adoption of mass spectrometry should lower false alarms rates often associated with ETD due to the higher resolution of the core technology.[citation needed] It also uses a non-radioactive ionization method generally secondary electrospray ionization (SESI-MS). Primarily used in desktop ETD systems, mass spectrometry can be miniaturized for handheld ETD.

Photoacoustic spectroscopy

A laser beam tuned to a specific absorption line of a target gas passes through a gas sample. When the laser intensity is modulated, the gas molecules absorb the light and release thermal energy via non-radiative relaxation, creating periodic acoustic pressure waves.The acoustic pressure waves force the prongs of the quartz tuning fork (QTF) to vibrate. Because quartz is piezoelectric, this mechanical deflection converts directly into an electrical voltage or current signal.

References

  1. ↑ Marshall, Maurice; Oxley, Jimmie (2009). ASPECTS OF EXPLOSIVES DETECTION.
  2. ↑ Rabbany, Sina Y., et al. "Trace detection of explosives using a membrane-based displacement immunoassay." Journal of immunological methods 246.1-2 (2000): 69-77.
  3. ↑ Mirasoli, Mara, et al. "Development of a chemiluminescence-based quantitative lateral flow immunoassay for on-field detection of 2, 4, 6-trinitrotoluene." Analytica chimica acta 721 (2012): 167-172.
  4. ↑ Romolo, Francesco Saverio, et al. "Field detection capability of immunochemical assays during criminal investigations involving the use of TNT." Forensic science international 246 (2015): 25-30.
  5. ↑ Narang, Upvan, Paul R. Gauger, and Frances S. Ligler. "A displacement flow immunosensor for explosive detection using microcapillaries." Analytical Chemistry 69.14 (1997): 2779-2785.
  6. ↑ Girotti, Stefano, et al. "Development of a chemiluminescent ELISA and a colloidal gold-based LFIA for TNT detection." Analytical and bioanalytical chemistry 396.2 (2010): 687-695.
  7. ↑ Fetterolf, D. D., J. L. Mudd, and K. Teten. "An enzyme-linked immunosorbent assay (ELISA) for trinitrotoluene (TNT) residue on hands." Journal of forensic sciences 36.2 (1991): 343-349.
  8. ↑ Wekalao, Jacob, et al. "Graphene and gold metasurface-based terahertz surface plasmon resonance sensor for explosive detection." Plasmonics 19.6 (2024): 3131-3145.
  9. ↑ Wang, Jin, et al. "Highly selective rational design of peptide-based surface plasmon resonance sensor for direct determination of 2, 4, 6-trinitrotoluene (TNT) explosive." Sensors and Actuators B: Chemical 264 (2018): 279-284.
  10. ↑ Brulé, Thibault, et al. "A field-deployed surface plasmon resonance (SPR) sensor for RDX quantification in environmental waters." Analyst 142.12 (2017): 2161-2168.
  11. ↑ Tam, Maggie, and Herbert H. Hill. "Secondary electrospray ionization-ion mobility spectrometry for explosive vapor detection." Analytical chemistry 76.10 (2004): 2741-2747.
  12. ↑ Itozaki, Hideo, and Go Ota. "Nuclear quadrupole resonance for explosive detection." International journal on smart sensing and intelligent systems 1.3 (2008): 705-715.
  13. ↑ Cardona, Lorena, Jovani Jiménez, and Nelson Vanegas. "Nuclear quadrupole resonance for explosive detection." Ingeniare. Revista chilena de ingeniería 23.3 (2015): 458-472.
  14. ↑ Pesenti, Alessandra, et al. "Coupling paper-based microfluidics and lab on a chip technologies for confirmatory analysis of trinitro aromatic explosives." Analytical chemistry 86.10 (2014): 4707-4714.
  15. ↑ Taranto, Vitor, et al. "The analysis of nitrate explosive vapour samples using Lab-on-a-chip instrumentation." Journal of Chromatography A 1602 (2019): 467-473.
  16. ↑ Thiesan, Lisa, et al. "Survey of commercially available explosives detection technologies and equipment 2004." Report for NIJ grant 96-MU-MU-K011 (2004).
  17. ↑ Gottfried, Jennifer L., et al. "Laser-induced breakdown spectroscopy for detection of explosives residues: a review of recent advances, challenges, and future prospects." Analytical and bioanalytical chemistry 395.2 (2009): 283-300.
  18. ↑ Jiménez, A. M., and M. J. Navas. "Chemiluminescence detection systems for the analysis of explosives." Journal of hazardous materials 106.1 (2004): 1-8.
  19. ↑ Jiang, Xinghe, et al. "Detection and imaging of chemicals and hidden explosives using terahertz time-domain spectroscopy and deep learning." Light: Science & Applications 15.1 (2026): 80.
  20. ↑ Shen, Y. C., et al. "Detection and identification of explosives using terahertz pulsed spectroscopic imaging." Applied physics letters 86.24 (2005).
  21. ↑ Wehmeyer, Alex L. The detection of explosives using an inertial electrostatic confinement DD fusion device. Diss. University of Wisconsin--Madison, 2005.
  22. ↑ Nunes, W. V., et al. "Explosives detection using prompt-gamma neutron activation and neural networks." Applied radiation and isotopes 56.6 (2002): 937-943.
  23. 1 2 Cumming, Colin; Fisher, Mark; Sikes, John (2004-01-01). Gardner, Julian W.; Yinon, Jehuda (eds.). Electronic Noses & Sensors for the Detection of Explosives. NATO Science Series II: Mathematics, Physics and Chemistry. Springer Netherlands. pp. 53–69. doi:10.1007/1-4020-2319-7_4. ISBN 9781402023170.
  24. ↑ Thomas, Samuel W.; Joly, Guy D.; Swager, Timothy M. (2007-04-01). "Chemical Sensors Based on Amplifying Fluorescent Conjugated Polymers". Chemical Reviews. 107 (4): 1339–1386. doi:10.1021/cr0501339. ISSN 0009-2665. PMID 17385926.
  25. ↑ Martínez-Lozano, Pablo; Rus, Juan; Fernández de la Mora, Gonzalo; Hernández, Marta; Fernández de la Mora, Juan (February 2009). "Secondary electrospray ionization (SESI) of ambient vapors for explosive detection at concentrations below parts per trillion". Journal of the American Society for Mass Spectrometry. 20 (2): 287–294. Bibcode:2009JASMS..20..287M. doi:10.1016/j.jasms.2008.10.006. ISSN 1044-0305. PMID 19013080.
  26. ↑ Tam, Maggie; Hill, Herbert H. (May 2004). "Secondary Electrospray Ionization-Ion Mobility Spectrometry for Explosive Vapor Detection". Analytical Chemistry. 76 (10): 2741–2747. Bibcode:2004AnaCh..76.2741T. doi:10.1021/ac0354591. ISSN 0003-2700. PMID 15144183.
  27. ↑ Aernecke, Matthew J.; Mendum, Ted; Geurtsen, Geoff; Ostrinskaya, Alla; Kunz, Roderick R. (2015-11-25). "Vapor Pressure of Hexamethylene Triperoxide Diamine (HMTD) Estimated Using Secondary Electrospray Ionization Mass Spectrometry". The Journal of Physical Chemistry A. 119 (47): 11514–11522. Bibcode:2015JPCA..11911514A. doi:10.1021/acs.jpca.5b08929. ISSN 1089-5639. PMID 26505487.
  28. ↑ Viola, Roberto, et al. "Quartz enhanced photoacoustic spectroscopy for detection of improvised explosive devices and precursors." Advances in Optical Technologies 2016.1 (2016): 5757361.
  29. ↑ Sharma, Ramesh, et al. "Ultrasonic standoff photoacoustic sensor for the detection of explosive and hazardous molecules." Defence science journal 68.4 (2018): 401-405.