Fundamental resolution equation

The fundamental resolution equation or Purnell equation is used in chromatography to help relate adjustable chromatographic parameters to resolution.

Equation

Rs=(N4)(α−1α)(k'21+k'2)

where,

N = Number of theoretical plates

α = Selectivity Term = k'2k'1

The N4 term is the column factor, the α−1α term is the thermodynamic factor, and the k'21+k'2 term is the retention factor. The 3 factors are not completely independent, but can be treated as such.

Intervention

To increase resolution of two peaks on a chromatogram, one of the three terms of the equation need to be modified.

Resolution

The fundamental resolution equation is derived as follows:

For two closely spaced peaks, ω1=ω2 , and σ1=σ2,

so,

Rs=tr2−tr1ω2=tr2−tr14σ2

Where tr1 and tr2 are the retention times of two separate peaks.

Since N=(tr2σ2)2 , then σ=tr2N

Using substitution, RS=N(tr2−tr14tr2)=(N4)(1−tr1tr2).

Now using the following equations and solving for tr1 and tr2

k'1=tr1−t0t0;tr1=t0(k'1+1)

k'2=tr2−t0t0;tr2=t0(k'2+1)

Substituting again and you get:

Rs=(N4)(1−k'1+1k'2+1)=(N4)(k'2−k'11+k'2)

And finally substituting once more α=k'2/k'1 and you get the Fundamental Resolution Equation:

Rs=(N4)(α−1α)(k'21+k'2)