Specific detectivity

Specific detectivity, or D*, for a photodetector is a figure of merit used to characterize performance, equal to the reciprocal of noise-equivalent power (NEP), normalized per square root of the sensor's area and frequency bandwidth (reciprocal of twice the integration time).

Specific detectivity is given by D∗=ANEP, where A is the area of the photosensitive region of the detector, and NEP the noise equivalent power (in units of W/Hz). It is commonly expressed in Jones units (cm⋅Hz/W) in honor of Robert Clark Jones who originally defined it.

Given that noise-equivalent power can be expressed as a function of the responsivity ℜ (in units of A/W or V/W) and the noise spectral density Sn (in units of A/Hz1/2 or V/Hz1/2) as NEP=Snℜ, it is common to see the specific detectivity expressed as D∗=ℜ⋅ASn.

It is often useful to express the specific detectivity in terms of relative noise levels present in the device. A common expression is given below.

D∗=qληhc[4kTR0A+2q2ηΦb]−1/2

With q as the electronic charge, λ is the wavelength of interest, h is the Planck constant, c is the speed of light, k is the Boltzmann constant, T is the temperature of the detector, R0A is the zero-bias dynamic resistance area product (often measured experimentally, but also expressible in noise level assumptions), η is the quantum efficiency of the device, and Φb is the total flux of the source (often a blackbody) in photons/s/cm2.

Detectivity measurement

Detectivity can be measured from a suitable optical setup using known parameters. You will need a known light source with known irradiance at a given standoff distance. The incoming light source will be chopped at a certain frequency, and then each wavelength will be integrated over a given time constant over a given number of frames.

In detail, we compute the detection bandwidth Δf directly from the integration time constant tc.

Δf=12tc

Next, an average signal and rms noise needs to be measured from a set of N frames. This is done either directly by the instrument, or done as post-processing.

Signalavg=1N(∑iNSignali)
Noiserms=1N∑iN(Signali−Signalavg)2

with the NEP=NoisermsΔf. Now, the computation of the radiance H in W/sr/cm2 must be computed where cm2 is the emitting area. Next, emitting area must be converted into a projected area and the solid angle; this product is often called the etendue. This step can be obviated by the use of a calibrated source, where the exact number of photons/s/cm2 is known at the detector. If this is unknown, it can be estimated using the black-body radiation equation, detector active area Ad and the etendue. This ultimately converts the outgoing radiance of the black body in W/sr/cm2 of emitting area into one of W observed on the detector.

The broad-band responsivity, is then just the signal weighted by this wattage.

R=SignalavgHG=Signalavg∫dHdAddΩBB,

where

From this metric noise-equivalent power can be computed by taking the noise level over the responsivity.

NEP=NoisermsRΔf=NoisermsSignalavgHGΔf

Similarly, noise-equivalent irradiance can be computed using the responsivity in units of photons/s/W instead of in units of the signal. Now, the detectivity is simply the noise-equivalent power normalized to the bandwidth and detector area.

D∗=AdNEP=ΔfAdHGSignalavgNoiserms

See also

References

  1. ↑ Rogalski, A. (2012-01-01). "History of infrared detectors". Opto-Electronics Review. 20 (3). doi:10.2478/s11772-012-0037-7. ISSN 1896-3757.
  2. ↑ R. C. Jones, "Quantum efficiency of photoconductors," Proc. IRIS 2, 9 (1957)
  3. ↑ R. C. Jones, "Proposal of the detectivity D** for detectors limited by radiation noise," J. Opt. Soc. Am. 50, 1058 (1960), doi:10.1364/JOSA.50.001058)

Public Domain This article incorporates public domain material from Federal Standard 1037C. General Services Administration. Archived from the original on 2022-01-22.